Generate an accurate turbidity reading quickly with the Thermo Scientific™ Orion™ AQUAfast AQ3010 Turbidity Meter

Generate an accurate turbidity reading quickly with the Thermo Scientific™ Orion™ AQUAfast AQ3010 Turbidity Meter. Perfect for the plant, field or lab, this simple yet powerful turbidity meter offers an impressive array of features. Use the Infrared (IR) LED light source to comply with ISO 7027 requirements and reliably measure turbidity with auto-ranging values from 0 to 1000 NTU. This IP67 waterproof meter offers a long battery life averaging about 3000 tests.

Features:
  • Easy to read results from the large, bright LCD display in a variety of work environments
  • Autoranging from 0 to 1000 NTUs
  • Uses Nephelometric principle with 90° detection
  • Meets ISO 7027 and offers consistent light source for greater accuracy
  • No mixing or diluting with EPA-approved polymer bead primary standards for faster calibration

Ordering Alerts:

Replacement products include: Calibration Standards (AC301S), Silicone Oil and Cloth AC3SIL, and Vial AC3V25. Other accessories include Cleaning Brush AC3CBR and Formazin Turbidity Standard (4000 NTU, 475mL) AC45FZ.

TRUSTED_SUSTAINABILITY

Spécification

Type Turbidity Meter
Light Source Infrared LED
Display Type LCD
Response Time <6 sec.
Description Turbidity meter kit
Key Functions Power, Cal, Scroll Up, Scroll Down, Read/Enter
IP Rating IP67
Precision ±2% of reading ±1 LSD for 0-500 NTU, ±3% of reading ±1 LSD for 501-1000 NTU
Resolution 0.01 NTU (0.01 to 19.99 NTU), 0.1 NTU (20.0 to 99.9 NTU), 1 NTU (100 to 1000 NTU)
Turbidity 0.01 to 19.99 NTU with 0.01 NTU resolution, 20.0 to 99.9 NTU with 0.1 NTU resolution, 100-1000 NTU with 1 NTU resolution
Battery Life 3000 Tests
Certifications/Compliance CE, TUV 3-in-1 for CSA-US, FCC part 15
ISO Accepted Methods ISO 7027
Temperature (Metric) Operating 0°C to 50°C
Afficher plus Afficher moins

Disclaimer to internal research use restriction: Any restrictions on the purchaser to utilize the product(s) for internal research purposes only does not apply to the product(s) on the web pages associated with this statement.

Correction du contenu d'un produit

Veuillez fournir vos retours sur le contenu du produit en remplissant le formulaire ci-dessous.

Nom du produit

En cliquant sur Soumettre, vous reconnaissez que vous pouvez être contacté par Fisher Scientific au sujet des informations que vous avez fournies dans ce formulaire. Nous ne partagerons pas vos informations à d'autres fins. Toutes les informations de contact fournies seront également conservées conformément à notre politique de confidentialité. Politique de confidentialité.

Merci de nous aider à améliorer notre site web. Votre commentaire a été soumis